Detail E+ H Ultrasonic Prosonic M FMU40
Quick and simple commissioning via menu-guided onsite operation with four-line plain text display; 7 languages selectable
Envelope curves on the on-site display for simple diagnosis
Easy remote operation, diagnosis and measuring point documentation with the supplied FieldCare operating program.
Suitable for explosion hazardous areas ( Gas-Ex, Dust-Ex)
Linearisation function ( up to 32 points) for conversion of the measured value into any unit of length, volume or flow rate
Non-contact measurement method minimizes service requirements
optional remote display and operation ( up to 20 m ( 66 ft) from transmitter)
Integrated temperature sensor for automatic correction of the temperature dependent sound velocity
Field of application
Continuous, non-contact level measurement in fluids, pastes, sullages and coarse bulk materials
Flow measurement in open channels and measuring weirs
System integration via:
HART ( standard) , 4 ... 20 mA
PROFIBUS PA
FOUNDATION Fieldbus
Maximum measuring range
5 m in fluids
2m in bulk materials
For other measuring ranges:
FMU41:
8m in fluids
3, 5 m in bulk materials
FMU42:
10 m in fluids
5 m in bulk materials
FMU43:
15 m in fluids
7 m in bulk materials
FMU44:
20 m in fluids
10 m in bulk materials
Tampilkan Lebih Banyak